    33**ZOOM:** https://imperial-ac-uk.zoom.us/j/98220889714?pwd=SM90vOF7BKSXoU3mD9q7OwBQo4IVB3.1
    44
    5 == Agenda ==
     5Attended: CW, KL, HO, AFr, MH, ND, CD, BT
     6
     7== Meeting Notes ==
    68
    791. Minutes and actions:
     
    1719 - **Done** Waiting
    1820- **RW:** Add lip to RCF holder
    19  - **Stands**
     21 - **Done**
    2022- **KL/CD:** Get RCF cut
    2123 - **Done**
    2224
    23252. Strathclyde: BT, ED, RW \\
    24 - Laser problem explanation
    25 - Quad focussing update
     26- Beamline update:
     27 - Missing pulses were traced to a failed power supply. New power supply gives an increase in energy. The increase in power was sufficient to move one of the system outside its linear range and compromise operation. There are ND filters in the system to correct for this and once the problem was identified and located, the level of filtering was corrected and the team were able to produce a good focal spot at close on Tuesday. On Wednesday, the quads were placed in the beamline and we saw proton focus. Today (Thursday), the quads were moved as close to the source as possible, and an increase in proton flux was seen. A z-scan has also been performed, waiting for resutls.
    2628- Dose Rate Update:
    27  - Tony results
    28  - Calvin Results
     29 - [raw-attachment: Dose_Rate_Calc.pptx Slides]
     30 - Calvin's sim produced ~1Gy. The RCF dose calculated by Nick peaked at 7Gy in the distribution. Tony produced a number of 2000Gy. Believe this must be a normalisation issue. Possibly a confusion of N vs E compared with dN/dE vs E
     31 - Clarify the problem to Tony so that he can calculate a more accurate dose per pulse
     32 - Will always have an issue with background subtraction that can cause a large variation in the calculated dose
     33 - Could remove some uncertainty by many RCF shots
     34 - Clear we need a faster way to calculate dose from the RCF
    2935
    30363. Simulation CD, DA, AFr: \\
    31 - Calvin optimisation Results
     37- Results indicated quads are best placed as close to the source as possible and as close to each other as possible. The gold foil should be at the tube entrance, and the cell dish should be as close to the vacuum window as possible. The importance of these reduces in the order of them being stated here.
     38- However, fear with the statistics used in the optimisation causing the noise to dominate and uniformity being lost to noise.
     39- Can improve the sims by reducing the bin sizes.
     40 - [raw-attachment: Optimal_gold_setup.pptx Slides]
    3241- Alfredo simulation results
    33 - Replacement for gold foil discussion
     42 - Indicated that a run with higher stats still looks uniform
     43- Highlighted an important issue, that in sim we can optimise number of protons against number of bins, there is a real world limit to these values in the number of protons per pulse and the area of a cell nucleus (110um^2).
     44 - Calculate the average proton to proton distance in simulation.
    3445
    35464. BioPrep EM, JP: \\
    36 - Incubator update
    37 - Consumables ordered
    38 - Tick off other items on checklist
     47- Tick off items on checklist
     48 - Incubator update - Double
     49 - Consumables ordered - Waiting for delivery
     50 - Waste stream - Sorted
     51 - Microscope - camera not needed
     52 - Hoods serviced - Check anything else needs doing to them? No
     53 - Ice machine - Sorted
    3954- Plan for staining finalised
    4055- Marie Boyd update
     
    4661- Real-time beam diagnostic update
    4762 - Lanex with a hole ready to go
     63 - Lanex also placed behind the cell dish
    4864
    49656. DoNM
    50 - Next Week: 30th Oct
     66- Next Week: 13th Nov
    5167
    52687. AoB