Changes between Version 6 and Version 7 of Research/Instrumentation/Meetings/2018-09-21-SmartPhantom


Ignore:
Timestamp:
Sep 25, 2018, 12:09:55 PM (6 years ago)
Author:
longkr
Comment:

--

Legend:

Unmodified
Added
Removed
Modified
  • Research/Instrumentation/Meetings/2018-09-21-SmartPhantom

    v6 v7  
    2626  * Smart phantom: concept and discussion: AK/All
    2727   * We '''agreed''' that the instrumented (smart) phantom that can deliver a measurement of LET/dose shot-by-shot was a valuable contribution.  So, we pursue the development of the sci-fi detector interleaved with the sample flasks.
    28    * AK presented the [Simulations.pdf:slides].  Points noted:
     28   * AK presented the [[https://ccap.hep.ph.ic.ac.uk/trac/attachment/wiki/Research/Instrumentation/Meetings/2018-09-21-SmartPhantom/Simulations.pdf:slides]].  Points noted:
    2929    * Two proton energy ranges to be considered: 130--180 MeV; 60--120MeV.  Proton energy spread ~1.5%.
    3030    * Dose rates over flask surface: 0.5, 1, 2, 4 and 6 Gy.  Delivered in spots of dimension 0.3x0.3 cm^2.  Each 'spot' has a fraction of the total dose, so that the dose to the flask adds up to the numbers given above.